Parameter
|
Value
|
3Â÷¼ö
|
Resistivity *
|
18.2 M§Ù .cm @ 25 ¡ÆC
|
TOC **
|
¡Â5 ppb (ug/L)
|
Pyrogens (endotoxins)
|
< 0.001 Eu/mL (pyrogen-free) after BioPak
|
RNases
|
< 0.01 ng/mL (RNase-free) after BioPak
|
DNases
|
< 4 pg/mL (DNase-free) after BioPak
|
Microorganisms
|
< 0.1 cfu/ml after Millipak or BioPak
|
Particulates > 0.22 um
|
< 1 particulate / mL after Millipak Express 40
|
Flow rate
|
Adjustable from trigger: from 50 mL /min up to 2 L/min
|
* Resistivity can be displayed temperature-compensated at 25 ¡ÆC or non-temperature-compensated as required by USP
** TOC spec - Test conditions: Milli-Q Direct system equipped with Q-Gard T1 pack and Quantum TEX cartridge. Product water quality
may vary due to local feed water conditions.
|
3Â÷¼ödelivery
|
Manual dispense flow rate
|
Adjustable between 50 and 2000 mL/min
|
Automatic dispense volume
|
100 mL, then 250 mL to 5 L by 250 mL increments, then 5 L to 60 L by 1 L increments
|
Volumetric dispense accuracy
|
3 % for volumes between 250 mL and 60 L
|
Volumetric dispense dispersion
|
CV < 3% for volumes between 250 mL and 60 L
|
1Â÷¼ö
|
Ions rejection
|
97 to 98% with new RO cartridge
|
Organics Rejection
|
> 99 % for MW > 200 Dalton
|
Particulates & Bacteria Rejection
|
> 99%
|
1Â÷¼ö production and delivery
|
Production Flow Rate
|
8 L/hour (Milli-Q Direct 8)
|
16 L/hour (Milli-Q Direct 16)
|
Delivery Flow rate
|
From tap: up to 2.5 L/min
|
From optional pump: up to 15 L/min at 1 bar
|