MET ONE 2400 SPECIFICATION
Size Channel(um) : 0.3um , 0.5um , 1.0um , 3.0um , 5.0um , 10.0um (6channel)
Flow Rate : 1.0cfm (28.3L/min)
Flow Control : Electronic meter with rear panel adjustment
Display : Red LED. 7 digits
Light Source : Laser Diode
Outputs : Built-in printer : RS232/RS485 serial interface
Sample and Hold Times : 1 second to 24 hours
Alarm Counts : 0 to 9,999,999
Location Labels : 0 to 999. appear on printout
Count Data : Total counts, counts/ft3 , counts/m3
Data Sorage : 400 samples
Coincidence Loss : Less than 5% at 400,000 counts/ft3
Power : 100VAC, 115VAC or 230VAC : 110 watts
Size : 11.2'w * 6.0'h *18"d(28.4 x 15.2 x 45.7cm)
Weight : 24 lbs(10.9kg)
Environment : Operaing 55 to 84 'F (12 to 29'C).
20 to 95% relative humidity. non-condensing
Storage -10 to 160'F(-23'C to 70'C).
Warranty : 1 year(instrument) 3 year (Laser Diode)
In addition to laser reliability, Met One Models 2400 and 2408 are easy-to-operate
workhorses. Plus, breakthrough Long Life Laser technology extends the average service
life of the instrumentation's laser to more than ten years for either model. A quiet,
carbon-free pump provides steady vacuum power at 1.0 cfm (28.3 L/min). The built-in
printer allows either immediate or delayed printing of results. Using Particle Vision?
PortAll¢â software, data can be downloaded at the operator¡¯s convenience from the
400-record buffer and displayed as a graph or a table in the built-in spreadsheet.
The bright LED display makes reading the data easy. The display shows
particle counts in either cumulative or differential modes, as well as the values from
the optional probes for humidity, temperature and air velocity. If your task includes
scanning filters, these counters can be ordered with a filter-scanning fan-shaped
probe that makes this task go more quickly and accurately. The electronic version
provides an audible and visual signal at the probe to help detect filter leaks.
For a fixed system, the Model 2400 or Model 2408 can be coupled with the
Model 2432 manifold, allowing sequential sampling of up to 32 sample points. This
combination can be an independent system or part of a computer-based FMS system.
The Model 2400 and Model 2408 offer reliability, versatility and ease-ofoperation
in a lightweight package. If your requirements are limited to detection of
0.5 micron and larger particles—choose the Model 2408. If you need sensitivity at
0.3 micron, then choose the Model 2400. Either choice will give you solid
performance in a quiet and lightweight package.
Lit. No. 5223
F10 Printed in USA
© Beckman Coulter Life Sciences 2013. All rights reserved.
In the interest of improving and updating its equipment, Beckman Coulter reserves the right to alter specifications to equipment at any time.